Wireless Radchex Green Meter (For Film/Screen)
Specifications
X-ray energy dependence:
Digital range: Minimum CRLU Rate: Maximum CRLU Rate: Power On/Off: Controls: Functions: Power requirements: Typical battery life between charging: Operating environment: Electronic Cassette: Weight: |
Simulates relative light output of Green Rare Earth Screens within +/- 3% over kVp range of 60 to 120 kVp and a patient equivalent thickness range of 5 to 35 cm (within specified operating rates)
Green Light Units; GNLU (AEC#); 0 to 655.35 1.50/sec (approx 0.15 mR/sec entrance exposure rate) 2500/sec (approx 250 mR/sec entrance exposure rate) Manual switch Software Reset Control Measures GNLU (AEC#); Calculates Optical Density Value from GNLU Built in NiMH rechargeable battery pack (9.6V) Greater then 15 hrs 59 degrees to 95 degrees F (15 to 35 degrees C) Dimensions 10” x 12” x 0.5” (24 cm x 30 cm x 1.3 cm) 1.8 kg (3.9 lb) |